Calibration of Disintegration Test Apparatus:
The disintegration test is one of the test parameters to measure the time of tablet to disintegrate in a defined medium.
There are 5 parameters for calibration of disintegration test apparatus.
Disintegration test calibration parameters are:
- Bath Sensor
- External Probe
- Timer
- Number of Stroke per minute
- Sieve Integrity Check
1.Bath Sensor:
- The temperature should be set to 37.5°c with the Temp key and save it.
- Go back to the main menu.
- Use purified water as a medium.
- Once the temperature reached 37.5°c, immerse the calibrated thermometer into the tank and wait for 5 minutes to stabilize the temperature.
- Record the temperature shown by the instrument and measured by a calibrated thermometer in the record book.
Acceptance Criteria:
Acceptance criteria for bath sensor is, the difference between temperature shown by the instrument and measured by a calibrated thermometer should not be deviated by ± 2.0°c from the set temperature.
Note: Additional temperature points can be performed according to requirements.
2. External Probe:
- Set the bath temperature at 37.0°c by using the Temp key and save it.
- Go to the main menu.
- Use purified water as a medium.
- Once the temperature reached 37.5°c, immerse the calibrated thermometer into the tank and wait for 5 minutes to stabilize the temperature.
- Record the temperature shown by the instrument and measured by a calibrated thermometer in the record book.
Acceptance Criteria:
Acceptance criteria for bath sensor is, the difference between temperature shown by the instrument and measured by a calibrated thermometer should not be deviated by ± 2.0°c from the set temperature.
Note: Additional temperature points can be performed according to requirements.
3. Timer:
- Prepare and save the method or protocol with 1 minute, 30 minutes, and 60 minutes. (As per requirement)
- Use purified water as a medium.
- Select a method or protocol with 1 minute in the individual mode for Basket A.
- Press the ‘RUN/HALT’ key of Basket A.
- Start the instrument as well as a calibrated stopwatch at the same time.
- At end of set time, stop the stopwatch immediately.
- Continue the procedure for a second 30 minute and a second 60 minute temperature point.
- Repeat the procedure for Basket B and Basket C.
Acceptance Criteria:
The difference between set time and time shown by calibrated stopwatch as follows:
Set Time | Acceptance Criteria |
1 min | ± 5 sec |
30 min | ± 60 sec |
60 min | ± 60 sec |
4. Number of Strokes per Minutes:
- Prepare and save the method or protocol with a set time of 1 minute.
- Select a method or protocol for Basket A.
- Once the temperature is reached press the ‘RUN/HALT’ key of Basket A.
- Start instrument as well as the manual count of strokes at the same time.
- At end of set time, stop the manual count immediately.
- Record the number of strokes on the record book.
- Repeat the procedure for Basket B and Basket C.
Acceptance Criteria:
Acceptance criteria for the number of strokes is, the number of strokes of basket rack assembly should be 29 to 32 cycles per minute.
5. Sieve Integrity Check:
- Physical verification of each sieve for intactness, free of cracks, abrasions, etc.
Acceptance Criteria:
The sieve must be Intact, Free from cracks and abrasions.
These are the 5 parameters are used for the calibration of disintegration test apparatus in the pharmaceutical industry.
Also Read:
Disintegration test Interview Questions and Answers
Dissolution test Interview Questions and Answers
Karl Fischer Titration Interview Question and Answers
For Interview Preparation Refer: Interview
For interview preparation refer
YT channel: Pharmabeej