Calibration of Bursting Strength Tester in pharma

Calibration Of Bursting Strength Tester:

Bursting Strength tester is calibrated for 2 parameters.

  1. Calibration of Board Head, 2. Calibration of Paper Head

Calibration of Board Head:

Perform the calibration in five trials by using standard Aluminum Foil for Board testers and report the individual and average results in calibration record.

Acceptance Criteria:

The average standard burst value of the aluminum foil shall be as per range given in the certificate.

Calibration of Paper Head:

Perform the calibration in five trials by using standard aluminum foil for paper testers and report the individual and average results in calibration record.

Acceptance Criteria:

The average standard burst value of the aluminum foil shall be as per the range given in the certificate.

Frequency of Calibration:

Board tester and paper tester shall be calibrated in once in three months and as recommended by the service engineer after maintenance job.